| Sign In | Join Free | My carsrow.com | 
              
  | 
      
Brand Name : drager
Model Number : 6850645
Certification : CE
Place of Origin : Germany
MOQ : 1pcs
Price : as discussed
Payment Terms : T/T, Western Union
Supply Ability : 300pcs / week
Delivery Time : 2-3 work days
Packaging Details : 1pcs / box
color : white
material : plastic and metal
orignal code : 6850645
package : box
weight : 25g
Electrical Connector : old plated slip rings
Drager 6850645 medical oxygen sensor o2 cell / oxygen battery
  
 Technical Specifications
  
| Measuring range | 0-100% | |||
| Accuracy | +/-2% of full scale | |||
| Signal output | 10-14mV | |||
| Linearity | +/-2% of full scale | |||
| Response T90 | 13 sec | |||
| Temp Coefficient | compensated | |||
| Shelf Life | 6 months | |||
| Humidity Non-condensing | 0-99% RH | |||
| Operating Temp | 0 to 40℃ | |||
| Recommended storage | 0 to 25℃ | |||
| Expected Life | 60 months | |||
| Warrangty | 18 months | |||
| Electrical Connector | gold plated slip rings |  		|||
 
  
| Measurement Range: | 0-100 % oxygen | 
| Accuracy and Repeatability: | < 1 % vol. O2when calibrated at 100 % O2 | 
| Zero Ofset: | < 0.5 % vol. O2in 100% N2, applied 5 minutes | 
| Linearity Error: | < 3 % relative | 
| Cross Interference: | < 0.5 % vol. O2response to: | 
| 10 % CO2balance N2 | |
| 80% N2O balance N2 | |
| 7.5% Halothane balance N2 | |
| 7.5 % Isofurane balance N2 | |
| 7.5 % Enfurane balance N2 | |
| 9% Sevofurane balance N2 | |
| 20% Desfurane balance N2 | |
| Infuence of Humidity: | - 0.03 % rel. per % RH at 25C | 
| Infuence of Pressure: | proportional to change in oxygen partial | 
| pressure | |
| Infuence of Mechanical Shock: | < 1% relative after a fall from 1m | 
| Operating Temperature: | 0 to 50C | 
| Temperature Compensation: | built-in NTC compensation (see below) | 
| Efect of Temperature | between +5 C and +25 C: 3 % relative error | 
| Compensation (steady state): | between 0 C and +50 C: 8 % relative error | 
| Operating Humidity: | 0-99 % RH non-condensing | 
| Long Term Output Drift: | < 1 % vol oxygen per month | 
| typically < - 15 % relative over lifetime | |
| Storage Temperature: | -20 to +50 C | 
| Recommended Storage: | +5 to +15 C | 
| Recommended Load: | 10 kOhms | 
| Warm-Up Time: | < 30 minutes, after replacement of sensor | 
| Weight: | approximately 28 grams | 
| approximately 43 grams E61912 series | 
 
 
 
| 
                                             | 
                
                        6850645 Draeger Oxygen Sensor O2 Cell / Oxygen Battery Plastic Metal Material Images |